Analysis Note
For every lot exact values of particle size and standard deviation are determined with an accuracy of 0.001 µm. The size determination of these particles follows procedures described by National Institute of Standards and Technology (NIST, USA) respectively by the Community Bureau of Reference (BCR), esp.: Transmission electron microscopy (TEM) Scanning electron microscope Light microscopy Coulter™ counter with MDF-systemCalibration of equipment is carried out with NIST-respectively BCR-particle size standards.
Application
Used to calibrate, qualify and monitor the performance of particle size analyzers, surface scanning instruments and particle counters.Also used:in acoustic traps to study the effect of particle detection on trapping performance of lab on a chip systemsas a reference of known dimensions to characterize real-time imaging systemsas a calibration standard to validate the Airy Disc First Fringe (ADFF) analysis technique based on size measurements.
Features and Benefits
suitable for routine instrument calibration checks, testing and corrections available in 5 mL pack size as a neat sample
General description
Monodisperse polystyrene based micro particles (size: 12.0 µm) are a particle size standard, designed to identify the particle size distribution (PSD) profile of test samples.