Temperature:

  • Working-temperature range 5°C above ambient tem...


     Read more

    Your Price
    $8,474.50 EACH

Safety at all times: High-end functions for the protection of cell cultures, bacteria cultures or tissue cultures.

Temperature:

  • Working-temperature range 5°C above ambient temperature up to +50°C
  • Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
  • Setting accuracy temperature 0.1°C
  • Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
  • Temperature variation in chamber at + 37°C +/- 0.3 K

Humidity:

  • Humidity control (standard) Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element
  • limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
  • Setting accuracy humidity 0.5 % rh
  • Setting range active humidity control (with option K7) 40 to 97 % rh and rh-Off

Control of standard components:

  • CO2 control - Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic
  • fault indication, barometric pressure compensation
  • Adjustment range - 0 to 20% CO2
  • Setting accuracy CO2 0.1%
  • Variation in time +/- 0.2 % CO2
  • Adjustment range 1 to 20 % O2
  • Setting accuracy 0.1 % O2

Control technology:

  • ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition
  • TFT-colour displays
  • Language setting German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese
  • Function SetpointWAIT the process time does not start until the set temperature is reached
  • Adjustable parameters temperature (Celsius or Fahrenheit), CO2, programme time, time zones, summertime/wintertime

Communication:

  • Interface Ethernet LAN, USB
  • Documentation programme stored in case of power failure
  • Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety:

  • AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
  • Autodiagnostic system integral fault diagnostics for temperature and CO2
  • Alarm visual and acoustic

Capacity, L107
% CO20 – 20
Dims, w x d x h, mm719 x 591 x 850
Int Dims, w x d x h, mm560 x 400 x 480
Temp Range, °C18 – 50
Weight, kg75