Haier Biomedical HR40-IIB2 Classic Series Biosafety Cabinet (100% Exhaust); 40 Inch

Code: H-HR40-IIB2 D2-100

Professional partial air purification equipment, it is suitable for cell biology, microbiology, biomedicine, biosafety and other related laboratories. It provides the most essential protection and ...


 Read more

Your Price
POA
Price on Application

Professional partial air purification equipment, it is suitable for cell biology, microbiology, biomedicine, biosafety and other related laboratories. It provides the most essential protection and isolation equipment for biosafety.

  • Complete air tight protection
    • Its structural airtightness has been strictly tested; after being pressurized to 500pa and maintained for 30 minutes, the pressure inside was no lower than 480Pa, at the same time it is completely airtight after going through the integrity test of high efficient particulate air (HEPA) filter, making it safe to use
  • Safe airflow system
    • The operating area utilises an airflow disruption design and its front window has air curtain protection at the upper edge and in the surroundings, which can prevent cross over air flow between the inside and outside and thus preventing the diffusion ofthe
    • internal and external pollutants;
    • The cabinet professional imported air pressure sensors can monitor the pressure changes in the positive and negative pressure zones in real-time and the sound and light alarms are automatically triggered when airflow changes exceed 20%;
    • The cabinet working space is surrounded by four negative pressure zones, which can ensure that the internal pollution is in negative pressure or is surrounded by negative pressure zones so that the operators are protected from the pollutants;
    • The cabinet is manufactured with an internationally renowned brand moisture-proof and flame-retardant glass fiber high efficiency particulate air (HEPA) filter, which ensures a filtration efficiency of 99.99% or above for particles of 0.3 microns to protect the samples, personnel and environment;
    • The cabinets high-efficiency AC centrifugal fans combined with professional air flow distribution design to ensure lower noise and more uniform air flow;
    • The cabinet LCD screen displays in real-time the various parameters and the service life of the components, making the equipment operating conditions clearer at a glance;
    • The cabinet high intensity 254nm UV lamp enables the setting of the automatic on/off timer for the sterilisation intervals, reducing user waiting times;
    • The cabinet has a linkage interlock system of ultraviolet disinfection, fluorescent lamp, front window and fan: after the front window has been opened, the fan will turn on automatically and the UV lamp will turn off to prevent the dispersal of pollutants, preventing any personnel injuries;
    • The cabinet has variety of light and sound alarm functions: hardware fault alarm, operating parameters out-of-limit alarm, filter and UV lamp service life alarm and other alarm functions
  • Friendly design
    • The front operating surface of the cabinet is manufactured with a 10° inclination design, which is in accordance with ergonomic principles and more comfortable to operate for the user;
    • The cabinet underframe has adjustable feet with no bare thread, avoiding dust pollution;
    • The cabinet has universal casters to facilitate movement;
    • Water valve, air valve, anti-backflow air valve and other accessories are optional
ModelHR40-IIB2
Power Supply, V/Hz220-240/50/60
Power, VA1700
Power of Blower, WAC 115
Airflow Circulation100% exhaust
Main Filter Typical EfficiencyULPA, U15, 99.9995% @0.12µm
Exhaust Filter Typical EfficiencyHEPA, H14, 99.995% @0.3µm
Exhaust FilterSingle
Filter BrandAAF
Downflow Velocity, m/s0.28
Inflow Velocity, m/s0.55
Lighting Intensity, Lux≥1200
Sound Level, dB(A)64
Weight, kg252
Internal Dims, w x d x h, mm1167 x 610 x 680
External Dims, w x d x h, mm 1360 x 820 x 2330
Support Stand, mm680mm (standard), 680 - 900mm adjustable (optional)
Sash Opening, mm190 (max 490)
AlarmSound and flash
Closing Fan Alarm After Door OpeningY
Door Ajar AlarmY
Abnormal Inflow AlarmY
Abnormal Downflow AlarmY
Door Open More than Limit AlarmY
Filter Clogging AlarmY
Filter Damage AlarmY
Filter Lifetime Lack AlarmY
UV light Lifetime Lack AlarmY
UV LampY
Socket1 socket
Valve PortStandard with 2 valve port
Side Wall Service TapsStandard gas tap and vacuum tap, optional compressed air tap and water tap
Antimicrobial Powder CoatingOptional
316SS Work SurfaceOptional
Work SurfaceOne-piece
Touch ScreenN
Motorised WindowN
Foot SwitchN
ArmrestY
Exhaust Duct Connection KitsOptional
CertificationCE, TUV
This product has met the following criteria to qualify for the following awards: