| Model | HCP-80 | HCP-168 | HCP-258 |
| Chamber Volume, L/Cu.Ft | 80/2.8 | 170/6.0 | 258/9.1 |
| Interior Chamber Construction | 304 stainless steel |
| Exterior Chamber Construction | Cold-rolled steel powder coated |
| Access Port | / | 42mm diameter | 35mm diameter |
| Data Outputs | Remote alarm contacts, USB |
| Net/Gross Weight (approx) kg | 75/95 | 95/130 | 110/155 |
| Interior Dims, w x d x h, mm | 400 x 420 x 490 | 490 x 560 x 650 | 570 x 610 x 745 |
| Exterior Dims, w x d x h, mm | 625 x 684 x 735 | 714 x 812 x 887 | 794 x 867 x 985 |
| Shelf Dims, w x d, mm | 380 x 300 | 473 x 434 | 550 x 484 |
| No. of Shelves Standard/Maximum | 3/8 | 3/11 | 3/13 |
| Max. Load Per Shelf/Total Load, kg | 15/45 |
| Shelf Construction | Perforated, adjustable |
| Rated Voltage Power Supply, V/Hz | 220-240/50/60 |
| Nominal Consumption, kW (Steri-Run) | 0.08 (0.85) | 0.095 (1.3) | 0.12 (1.35) |
| Controller | Microprocessor |
| Display | 7″ LCD screen |
| CO2 Control Accuracy, % | 0.10% |
| CO2 Range, % | 0 - 20 |
| CO2 Alarm Range, % | ±0.5 |
| CO2 Inlet Pressure, Psi (Bar) | 12 - 17 (0.8 - 1.2) |
| CO2 Gas Purity, % | Min. 99.5 or medical quality |
| CO2 Inlet | 1/8″ hose (barbed) |
| CO2 Sensor | IR |
| CO2 Recovery Time* Min | 4 |
| CO2 Inlet Filter, µm | ﹤0.2 |
| High/Low Temperature Alarm | Y |
| Remote Alarm | Y |
| Sensor Error Alarm | Y |
| Excessive CO2 Concentration Alarm | Y |
| Water Shortage Reminder Alarm | Y |
| Door Ajar Alarm | Y |
| Temperature Control Accuracy, °C | 0.1 |
| Temperature Range, °C | Ambient +3 to 55 |
| Temperature Uniformity, °C | ±0.3 |
| Temperature Ambient Range, °C | 18 - 32 |
| Temperature Fluctuations, °C | ±0.1 |
| Temperature Sensor | 2*PT1000 |
| Temperature Recovery Time** Min | 4 |
| Sterilisation Cycle Temperature, °C | 180 dry heat sterilisation |
| Sterilisation Cycle Duration | Under 12 Hours |
| Humidity RH at 37°C, % | 93 ± 3 |
| Humidity Reservoir | Max. 1.75L/min 0.5L | Max. 3.5L/min 0.5L | Max. 5.5L/Min 0.5L |
| HEPA Filter | Optional | Optional | Optional |
| Pressure Reducing Valve | Optional | Optional | Optional |
| 4-20mA | Optional | Optional | Optional |
| The Cylinder Switch | Optional | Optional | Optional |
| Shelf | Optional | Optional | Optional |
| Water Tray | Optional | Optional | Optional |
| 3 Inner Door | N | Optional | N |
| 6 Inner Door | N | Optional | N |
| 8 Inner Door | N | N | Optional |
| Roller Base | Optional | Optional | Optional |
| Pure Copper Inner Liner | Optional | Optional | Optional |
| Pure Copper Shelf | Optional | Optional | Optional |
| Humidity Display | N | Optional | N |
| Oxygen Module | Optional | Optional | Optional |
| Electromagnetic Lock | N | Optional | N |
| Heightening Stand | Optional | Optional | Optional |
| IoT | Optional | Optional | Optional |
| Certification | CE | CE | CE |
| ** After 30s door opening, 98% from initial value, for CO2 not exceeding 5.2% |
| ** After 30s door opening, 98% from initial value, for temperature not exceeding 37°C |